三明市MedAccred _AC8121电子电缆和线束组件的审核标准 (95)

40.3.31 C- If crimp height testing is required by customer, is there evidence that test procedure is being followed?YES/NO/NA

40.3.32 C- If crimp force monitoring is required by customer, is there evidence that either crimp height or pull force testing was used to verify an acceptable crimp before referencing the crimp force monitor?YES/NO/NA

40.3.33 C- If crimp force monitoring is required by customer, is there evidence of compliance to the required crimp forces, in accordance with procedures/ customer requirements?YES/NO/NA

40.3.34 C- If crimp force monitoring is required by customer, is there evidence that test procedure is being followed?YES/NO/NA

40.3.35 C- If RF connector / coaxial shield pull force testing is required by customer, is there evidence that pull force test samples passed the requirement minimum pull forces, in accordance with procedures / customer requirements?YES/NO/NA

40.3.36 C- If RF connector / coaxial shield pull force testing is required by customer, are samples used for this test prevented from being used in deliverable product?YES/NO/NA

40.3.37 C- If RF connector / coaxial shield pull force testing is required by customer, is there evidence that test procedure is being followed?YES/NO/NA

40.3.38 C- If RF connector shield ferrule torsion testing is required by customer, is there evidence that test procedure is being followed?YES/NO/NA

40.3.39 C- If other customer defined mechanical testing is required, is there evidence that test procedure is being followed?YES/NO/NA

(0)

相关推荐